Recall Z-3279-2018 ยท LNH System, Nuclear Magnetic Resonance Imaging
GE Medical Systems, LLC
Discovery MR750w Nuclear Magnetic Resonance Imaging System labeled as the following product name sold under: Product Name (Model Number): A. 3.0T MR750 32CH (911000-3); B. 3.0T MR750W (5911000-3, 5481490-2, 5499460-7, 5499460-10, 5373011-3, 5352293, 5911000-3); C. 3.0T MR750W GEM (5499460-10, 5481490-2); D. 3.0T MR750W GEM ENAB (5911000-3); E. 3.0T MR750W GEM ENABLED (5352293, 5481490-2, 548
| Date | Jun 6, 2018 |
| Status | Terminated |
| Reason | GE Medical Systems has discovered a small area on the bore surface of the Discovery MR750w Nuclear Magnetic Resonance Imaging System, normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used. |
| Root cause | Nonconforming Material/Component |
| Action | GE Medical Systems notified customers on or about 06/06/2018 via a customer letter entitled "Urgent Medical Device Correction". The letter directed its customers to continue to use the system by following the existing Operator Manual Safety Information for proper patient padding and observation during scanning. Safety information included the following: WARNING: Place appropriate non-conductive padding between the patient and the bore wherever a portion of the body may come into contact with the magnet opening. CAUTION: Continuous patient observation and contact are required in all modes o |
| Quantity | 516 Units Total |
| Distribution | Worldwide Distribution - US Nationwide and in the states of AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, US, UT, VA, WA, WI, and WY. and the countries of Algeria, Argentina, Australia, Austria, Bahrain, Belar |
| FDA record | Open on fda.gov |