ClearanceBrief
Recall Z-2311-2018 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis zee floor, Material no. 10094135, for angiography and whole body radiographic/fluoroscopic procedures.

DateMay 1, 2018
StatusTerminated
ReasonAfter the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
Root causeDevice Design
ActionOn May 1, 2018, the firm notified customers via Urgent Medical Device Correction letters. Users were warned that the Large Display may not show an image after it returns from power save mode. The Large Display stays black without showing an error message although X-Ray is still possible. The problem occurs sporadically, and does not occur during an ongoing procedure. The firm stated that there should be no influence on the previous treatment of patients. Users were instructed to follow standard emergency procedures in the event of system failures. It is recommended to consider reinforcing em
Quantity14
DistributionWorldwide distribution. US nationwide including Puerto Rico, Australia, Austria, Bahrain, Belarus, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Luxembourg, Mexico, Myanmar, Netherlands, New
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141574ARTIS ZEE/ZEEGO SW VC21
Siemens Medical Solutions USA, Inc., Sep 5, 2014