Recall Z-2283-2018 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model 10849000

DateMay 1, 2018
StatusTerminated
ReasonIf the C-arm is manually moved in the longitudinal direction to a fully extended position, the system's safety stop mechanism may be activated. As a result, the C-arm stops the movement and can drop up to 20 mm (0.8 ). Under certain circumstances, the cover of the detector or collimator can touch and possibly injure the patient due to the C-arm drop. This could lead to a situation in which it is necessary to stop the clinical treatment or continue the treatment on an alternative system.
Root causeDevice Design
ActionOn May 1, 2018, the firm distributed Urgent Medical Device Correction letters to its affected customers. The letter instructed customers to follow standard emergency procedures they have in place in the event of system failures. The firm recommended considering reinforcing the emergency procedures until the update has been completed. As an immediate measure, operators were instructed to avoid driving the C-arm into a fully extended longitudinal position. The firm is currently developing a solution to eliminate the root cause of the problem. The solution will be distributed to all affected cu
Quantity39
DistributionHealthcare facilities in AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, NE, NJ, NY, OH, TN, TX, UT, and VA. Foreign distribution to Australia, Austria, Belgium, Denmark, France, Germany, Italy, Japan, Kuwait, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Ar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163286ARTIS pheno
Siemens Medical Solution USA, Inc., Mar 9, 2017