Recall Z-2335-2018 ยท MXC Recorder, Event, Implantable Cardiac, (Without Arrhythmia Detection)

St Jude Medical Inc.

Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.

DateApr 16, 2018
StatusTerminated
ReasonThe firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.
Root causeNonconforming Material/Component
ActionThe firm, Abbott, sent an "URGENT MEDICAL DEVICE CORRECTION" letters dated 05/18/2018, to its consignees notifying them that updated programmer software version 24.2.x will be made available over the next several weeks and will allow physicians to detect the presence of incorrect battery indicator prior to implant. Additionally, the programmer software will provide a mechanism to resolve the incorrect display for already implanted devices. If a low battery indicator is observed, customers were advised to contact firm technical services at 1-800-722-3774 to assist in confirmation and correction
Quantity0
DistributionU.S. Nationwide Distribution: PR, MI, NJ, IL, FL, AR, PA, KY, CA, TN, IN, AL, TX, AZ, SC, OK, VA, NH, NM, WA, CO, IA, WI, NV, UT, OH, NC, MO, DE, WV, MD, OR, NE, VT, KS, MA, IA, WY, CT, GA, MS, LA, DC, LA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163407Confirm Rx Insertable Cardiac Monitor (ICM) System
St. Jude Medical, Inc., Sep 29, 2017