Recall Z-2317-2018 ยท DTE Pulse-Generator, Pacemaker, External

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)

Single Chamber Temporary External Pacemaker, Model 53401

DateJun 29, 2018 (initiated Jun 6, 2018)
StatusTerminated
ReasonA subset of Medtronic Model 53401 Single Chamber External Pulse Generators (EPGs) may revert from programmed settings to nominal settings during use with a patient due to an unexpected Power on Reset (POR).
Root causeDevice Design
ActionMedtronic began notifying customers, via letter, on about 06/06/2018. Customers were instructed to verify if they possessed affected 53401 EPGs, if they did have affected EPGs to determine if the firmware was updated, to notify all appropriate personnel, to notify customers if the affected devices have been transferred, and to complete and return the Customer Confirmation Certificate. Instructions for determining if the EPG firmware has been updated include: 1) powering the unit on and viewing firmware version; 2) if the firmware version is 01.03.00 or greater, the EPG firmware has been up
Quantity1,108 devices
DistributionDomestic distribution to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV. International distribution to Armenia, Austria, Bangladesh, Belgium, China, Dominican Republ
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162054Medtronic Temporary External Pacemaker 53401
Medtronic, Inc., Oct 18, 2016