Recall Z-2340-2018 ยท DQK Computer, Diagnostic, Programmable

GE Healthcare Finland Oy

GE Healthcare CARESCAPE Monitor B650

DateMay 25, 2018
StatusOpen, Classified
ReasonWhen multiple CARESCAPE Monitor B650 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
Root causeSoftware design
ActionThe firm notified its consignees of the problem by letter on 05/25/2018. The letter stated the following safety instructions: "If a prolonged network overload occurs and CARESCAPE Monitor B650 on the network does not resume normal functionality in approximately 90 seconds, take the following actions: 1. Temporarily switch the monitor from central monitoring to local-only (bedside) monitoring: - Disconnect the LAN cables from IX/MC ports on backside of the monitor. The monitor will now be in local-only mode. - Keep the patient under close observation and continuous surveillance at t
Quantity69231 units
Distributionworldwide
FDA recordOpen on fda.gov