Recall Z-2745-2018 ยท LCP Assay, Glycosylated Hemoglobin

Tosoh Bioscience Inc

G8 Automated HPLC Analyzer: HLC-723G8-ST, 021560; and HLC-723G8-LA, 021674 Product Usage: The Tosoh Automated Glycohemoglobin Analyzer HLC-723G8 is intended for in vitro diagnostic use for the quantitative measurement of % hemoglobin A1c (HbA1c) (DCCT/NGSP) and mmol/mol hemoglobin A1c (IFCC) in whole blood specimens. This test is to be used as an aid in diagnosis of diabetes and as an aid in iden

DateJun 6, 2018
StatusTerminated
ReasonHbAE is known to interfere with the HbA1c assay on the current version of software, Ver. 5.23. Customers should exercise caution when reviewing chromatograms and ensure that Flag 43 is enabled on their device to avoid reporting invalid test results in the presence of HbAE.
Root causeSoftware design
ActionThe firm issued an Urgent Medical Device Recall letter to its affected customers on June 6, 2018. In the notice, Tosoh is currently investigating a resolution to this issue to eliminate interference from hemoglobin variants. In the interim, customers should carefully review the chromatography to determine whether or not a variant peak is detected and follow the recommendations included in the letter. Customers were asked to complete and return the attached Acknowledgement Form and the flag setting confirmation to Tosoh Bioscience, Inc., within 14-days of receiving this notification by FAX
Quantity665
DistributionWorldwide Distributed - US Nationwide. and the countries of Foreign distribution to Brazil, Caribbean, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Panama, Peru, and Uruguay.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100304ST AIA PACK HBA1C
Tosoh Bioscience, Inc., Aug 27, 2010