ClearanceBrief
Recall Z-2138-2018 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

12C

DateSep 16, 2016
StatusTerminated
ReasonIBA is initiating this recall to notify its customers that the use of non-IBA accessories and/or third party equipment or unauthorized modifications to an IBA medical device may pose a safety risk.
Root causeFinished device change control
ActionIBA mailed an Urgent Field Safety Notice to affected customers on 09/16/2016 to notify them of the risks associated with the use of this device to prevent serious injury/death. Customers were asked not to make any unauthorized modification to the IBA System or IBA accessories unless approved in writing or in advance by IBA. IBA will be updating their manuals to include these warnings.
QuantityS0 Units
DistributionUnits were delivered in the USA, Korea, Germany, France, Czech Republic, Italy, Poland and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132847I2C
Ion Beam Applications S.A., Nov 20, 2013