ClearanceBrief
Recall Z-2284-2018 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) Proteus 235, Beam Management System (PBS)

DateApr 11, 2016
StatusTerminated
ReasonIBA is conducting a voluntary recall to address a PTS (Proton Therapy System) software issue and to reduce the risk related to this issue.
Root causeSoftware Manufacturing/Software Deployment
ActionIBA mailed an Urgent Field Safety Notice to affected customers to inform them and prevent risk associated with this issue. IBA also recommends users always check for any discrepancy between the prescribed and delivered MUs on their clinical screen after an irradiation in which a pause occurred. IBA also recommends users always check for any discrepancy between the prescribed and delivered MUs on their clinical screen after an irradiation in which a pause occurred. A software upgrade will be deployed on impacted units to correct this issue.
Quantity2
DistributionUnits were distributed to Korea and Jacksonville, FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152224Proton Therapy System - Proteus 235 (Proteus One, Proteus Plus, Proteus TK2)
Ion Beam Application S.A., Aug 9, 2016