Recall Z-2279-2018 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)
Randox Laboratories Ltd.
Randox Liquid Assayed Specific Protein Control - Level 2 (SP Control 2), Cat. no. PS2683
| Date | Jun 13, 2018 (initiated May 22, 2018) |
| Status | Terminated |
| Reason | The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains. |
| Root cause | Under Investigation by firm |
| Action | On May 22, 2018, Randox issued an Urgent Field Safety Notice to affected distributors along with an Important Notice and updated IFU. The letter advised customers of the product issue. It further instructed distributors to do the following: 1. Discuss the contents of the notice with your Medical Director. 2. Place a copy of the important notice and updated IFU into any remaining stock. 3. Complete and return the vigilance response section of this form to technical.services@randox.com within five working days. If you have any questions after reviewing this information, you may contact the fir |
| Distribution | Distributed to accounts in CA, IN, KS, ME, and WV, and PR. Foreign distribution to Australia, Bangladesh, Belarus, Belgium, Brazil, Cambodia, Canada, Chile, China, Colombia, Cyprus, Denmark, Dutch Caribbean, Ecuador, Egypt, El Salvador, Ethiopia, Falkland Islands, Finland, France, Georgia, Germany, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K972608 | RANDOX ASSAYED SPECIFIC PROTEIN CONTROL Randox Laboratories, Ltd., Aug 4, 1997 |