ClearanceBrief
Recall Z-2254-2018 ยท BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Perseus A500 Anesthesia Machine; Cat. no. MK06000 Product V Product Usage: Intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressuresupported spontaneous breathing, and spontaneous breathing. Perseus is equipped with airway monitoring, gas measurement and device monitoring, O2 insufflation, and an anesthetic gas receiving syst

DateMay 24, 2018
StatusTerminated
ReasonThe Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
Root causeProcess change control
ActionOn May 24, 2018, the firm issued Urgent Medical Device Recall letters to affected end users (medical facilities) via certified mail. The letter identified the affected product, problem and actions to be taken. The recalling firm stated that they are currently manufacturing replacement parts for all potentially affected devices, which will be installed as part of the next service. The firm's Service dept. will contact end users to arrange the replacement Free of Charge In addition, end users were asked to complete and return a Customer Reply Card to acknowledge the receipt of this information
Quantity33
DistributionUS Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133886PERSEUS A500
Draeger Medical GmbH, Jun 20, 2014