ClearanceBrief
Recall Z-2252-2018 ยท BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Fabius GS MRI Anesthesia Machine; Cat. no. 8607300 Product Usage: Inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series machines are equipped with a compact breathing system; providing fresh gas decoupling, PEEP, and pressure limitatio

DateMay 24, 2018
StatusTerminated
ReasonThe Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
Root causeProcess change control
ActionOn May 24, 2018, the firm issued Urgent Medical Device Recall letters to affected end users (medical facilities) via certified mail. The letter identified the affected product, problem and actions to be taken. The recalling firm stated that they are currently manufacturing replacement parts for all potentially affected devices, which will be installed as part of the next service. The firm's Service dept. will contact end users to arrange the replacement Free of Charge In addition, end users were asked to complete and return a Customer Reply Card to acknowledge the receipt of this information
Quantity4
DistributionUS Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072884FABIUS MRI
Draeger Medical AG & Co. KG, Jul 30, 2008