Recall Z-2256-2018 ยท GEH Unit, Cryosurgical, Accessories

PENTAX of America Inc

C2 CryoBalloon Focal Catheter, REF FG-1009. C2 CryoBalloon Ablation System is used as a cryosurgical tool in the field of general surgery, specifically for endoscopic applications.

DateJun 5, 2018 (initiated Mar 30, 2018)
StatusTerminated
ReasonThe catheter alignment arrow printed onto the strain relief is not aligned with the detachment feature of the catheter connector, resulting in the potential for the catheter to be attached to the controller in the detach orientation. In this orientation, the catheter is not securely attached to the controller and may be partially (~1cm) ejected from controller.
Root causeVendor change control
ActionOn 3/30/2018, C2 Therapeutics issued URGENT FIELD CORRECTION notices to customers via certified mail, return receipt requested. Customers are advised to take the following Action: - Forward a copy of the notice and the enclosures to the department in which is referenced in the notice. - Inspect stock and quarantine products identified within the notice. - The end user of the affected products should complete and return the response form via email to RMA@c2therapeutics.com. The response form contains an RMA number to allow for the return of devices affected by this field action, and enclosed
Quantity57 units
DistributionDistribution US nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131523CRYOBALLOON FOCAL ABLATION SYSTEM
C2 Therapeutics, Aug 22, 2013