Recall Z-2266-2018 ยท CBK Ventilator, Continuous, Facility Use

Datex-Ohmeda, Inc.

CARESCAPE R860 ventilators with software version 10SP05 Product Usage: The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pr

DateJun 1, 2018 (initiated Dec 8, 2017)
StatusTerminated
ReasonPotential for the display processor to experience an unexpected failure.
Root causeSoftware change control
ActionOn December 12, 2017 GE Healthcare sent letters to their consignees stating that There is a potential for the display processor to experience an unexpected failure. If this failure occurs, the device screen will temporarily go blank and will then display an Error Code EC01 XX. The blank display results in loss of monitored data being displayed to the user. There have been no injuries reported as a result of this issue. You may continue to use the system. Device performance is not affected on devices where no such display failure occurs. In the event this issue occurs, an audible alarm will so
Quantity349 (US) + 2,051 (OUS)
DistributionWorldwide Distribution - USA (nationwide) Distribution to the states of : AL, AZ, CA, GA, ID, IL, IN, MS, NC, NJ, NK, NV, OH, OK, RI, TX, and WI., and to the countries of : Albania, Argentina, Australia, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142679CARESCAPE R860
Datex-Ohmeda, Inc., Jun 12, 2015