Recall Z-2531-2018 ยท LGX Table, Examination, Medical, Powered

Brewer Company, LLC

Access Hi-Low Exam Tables, Model 6501-XX

DateMar 29, 2018
StatusTerminated
ReasonOne lot, the wire harness was over-crimped on some of the connectors that go to the two vertical side switches on the exam table.
Root causeNonconforming Material/Component
ActionThe Brewer Company notified customers, via "MEDICAL DEVICE CORRECTION" letter, on about 03/21/2018. Instructions included to cease distribution of the product, coordinate with Brewer to schedule rework of the cable in affected tables as quickly as possible, and to notify Brewer of customers if the product was further distributed to coordinate rework of those tables. Additionally secondary controls were provided that the operator must press and continually hold the foot control pedal to operate the table, the foot control includes a stop button, depressing any pedal on the foot control wil
Quantity7 devices
DistributionDevices distributed in WI, TX, MN and IL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.