ClearanceBrief
Recall Z-2353-2018 ยท HAW Neurological Stereotaxic Instrument

FHC, Inc.

microTargeting" Lead Adapter for microTargeting" Drive, Product Number 66-CN-DB

DateMay 10, 2018
StatusTerminated
ReasonThe FHC Depth Stop Adapter may cause damage to the lead and stylet when the depth stop screw is over-tightened onto the lead. This damage can manifest as low impedances or a short circuit between one or multiple electrode pairs in the lead, and has resulted in the need to remove and replace the implanted lead intraoperative or in a follow-up procedure.
Root causeDevice Design
ActionUrgent Medical Device Recall notification letters were sent to customers on 5/10/18. The letters included the following instructions: FHC, Inc. Field Representatives will review your inventory and replace effected Depth Stop Adapters you have on hand with an improved design. In the interim, we advise that you do not over-tighten the Depth Stop Adapter screw. If low impedances or a short circuit is found, physicians should replace the lead prior to use and report the issue to FHC, Inc. or your local field rep.
Quantity2952 total
DistributionThe devices were distributed to the following US states: AL, AR, AZ, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The devices were distributed to the following foreign
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011992FHC MICROTARGETING DRIVE SYSTEM
FHC, Inc., Aug 14, 2001