ClearanceBrief
Recall Z-2227-2018 ยท GAX Tourniquet, Nonpneumatic

The Seaberg Company Inc

Part #: SAM XT-C, Model: Hi-Viz Orange or Civilian

DateMay 2, 2018
StatusTerminated
ReasonBased on internal testing, results indicated a possible failure of the stitches securing the buckle to the nylon belt could occur, posing a potential risk when used on a human patient to stop arterial blood flow. There is a remote chance that a serious injury and/or death could occur as a result of the failure of the device.
Root causeEmployee error
ActionOn 05/02/2018, the firm posted a press release notifying customers of a recall of all unused SAM XT Extremity Tourniquets (SAM XT) due to a possible failure of the stitches securing the buckle to the nylon belt that could pose a potential risk when used on a human patient to stop arterial blood flow. On 05/03/2018, recall notification forms were mailed to customers via overnight mail service.
DistributionUSA: CA, MD, OH, FL, ID, CO, CT, OR, AZ, IL, ME, MO, NY, SC, MI, AR, MT, CT, WA, TX, KY, NC, NJ, PA OUS: Australia, Canada, China, Columbia, Costa Rica, Europe (EU), Hong Kong, Iceland, India, Iraq, Israel, Japan, Kenya, Korea, Lebanon, Malaysia, Russia, Saudi Arabia, Singapore, South Africa, Thail
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.