Recall Z-2166-2018 ยท JSP Kit, Anaerobic Identification

BioMerieux SA

The VITEK 2 Anaerobic and Corynebacteria identification card (ANC) is intended for use with the VITEK 2 Systems for the automated identification of most clinically significant anaerobic organisms and Corynebatcerium species. The VITEK 2 ANC identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that cons

DateMar 23, 2018
StatusTerminated
ReasonA shipment to Guadeloupe suffered a temperature excursions up to 47.8C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioMrieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Root causeEnvironmental control
ActionThe firm, Biomerieux, sent an "Urgent product removal notice" letter on April 23 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Immediately acknowledge receipt (AR) of this FSCA. 2. Identify all countries and customers for which you are responsible that are impacted by this FSCA. 3. Define additional actions required (e.g. Translate customer letter, send to impacted countries/customers, etc.) 4. Distribute the customer letter to all customers that have received the products 5. Discard any impacte
Quantity135 (78 boxes of various product) in total
DistributionInternational distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.