Recall Z-2034-2018 ยท IKY Mattress, Flotation Therapy, Non-Powered

Joerns Healthcare

ProCair Mattress: (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 (c) PROCAIR PLUS 84" MATTRESS, MATERIAL NUMBER 222-6384 Product Usage: Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).

DateMay 3, 2018
StatusTerminated
ReasonJoerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.
Root causeEmployee error
ActionThe firm initiated the recall by letter on 05/04/2018. The letter requested the consignee identify potentially affected mattresses and contact Joerns Healthcare directly to schedule on-site rework of the unit. Distributors were directed to provide the recall information to their accounts.
Quantity46 units
DistributionWorldwide Distribution - US Nationwide in the states of UK, Canada, Holland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.