Recall Z-2034-2018 ยท IKY Mattress, Flotation Therapy, Non-Powered
Joerns Healthcare
ProCair Mattress: (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280 (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680 (c) PROCAIR PLUS 84" MATTRESS, MATERIAL NUMBER 222-6384 Product Usage: Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).
| Date | May 3, 2018 |
| Status | Terminated |
| Reason | Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells. |
| Root cause | Employee error |
| Action | The firm initiated the recall by letter on 05/04/2018. The letter requested the consignee identify potentially affected mattresses and contact Joerns Healthcare directly to schedule on-site rework of the unit. Distributors were directed to provide the recall information to their accounts. |
| Quantity | 46 units |
| Distribution | Worldwide Distribution - US Nationwide in the states of UK, Canada, Holland |
| FDA record | Open on fda.gov |