Recall Z-2033-2018 ยท IKY Mattress, Flotation Therapy, Non-Powered

Joerns Healthcare

P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA (b) PROMATT PLUS 84" MATRESS, NYLON, MATERIAL NUMBER PMP3684NZBA (c) PROMATT PLUS W/SIDEWALLS, 42X80, MATERIAL NUMBER PMS4280NZBA (d) PROMATT PLUS 36IN MATRESS, MATERIAL NUMBER PMP3680NZBA (e) PROMATT WITH SIDEWALLS ASSEMBLY, MATERIAL NUMBER PMS3680NZBA (f) PROMATT PLUS 80IN MATRESS, STRETCH, MATERIAL NUMBER PMP

DateMay 3, 2018
StatusTerminated
ReasonJoerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.
Root causeEmployee error
ActionThe firm initiated the recall by letter on 05/04/2018. The letter requested the consignee identify potentially affected mattresses and contact Joerns Healthcare directly to schedule on-site rework of the unit. Distributors were directed to provide the recall information to their accounts.
Quantity915 units
DistributionWorldwide Distribution - US Nationwide in the states of UK, Canada, Holland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.