Recall Z-2033-2018 ยท IKY Mattress, Flotation Therapy, Non-Powered
Joerns Healthcare
P.R.O. Matt: (a) PROMATT PLUS 36 X 76 MATRESS, NYLON, MATERIAL NUMBER PMP3676NZBA (b) PROMATT PLUS 84" MATRESS, NYLON, MATERIAL NUMBER PMP3684NZBA (c) PROMATT PLUS W/SIDEWALLS, 42X80, MATERIAL NUMBER PMS4280NZBA (d) PROMATT PLUS 36IN MATRESS, MATERIAL NUMBER PMP3680NZBA (e) PROMATT WITH SIDEWALLS ASSEMBLY, MATERIAL NUMBER PMS3680NZBA (f) PROMATT PLUS 80IN MATRESS, STRETCH, MATERIAL NUMBER PMP
| Date | May 3, 2018 |
| Status | Terminated |
| Reason | Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses. For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells. |
| Root cause | Employee error |
| Action | The firm initiated the recall by letter on 05/04/2018. The letter requested the consignee identify potentially affected mattresses and contact Joerns Healthcare directly to schedule on-site rework of the unit. Distributors were directed to provide the recall information to their accounts. |
| Quantity | 915 units |
| Distribution | Worldwide Distribution - US Nationwide in the states of UK, Canada, Holland |
| FDA record | Open on fda.gov |