ClearanceBrief
Recall Z-2164-2018 ยท LQL Gram Positive Identification Panel

BioMerieux SA

The VITEK 2 Gram-Positive identification card (GP) is intended for use with the VITEK 2 Systems for the automated identification of most significant Gram-positive organisms. The VITEK 2 GP identification card Is a single-use disposable. GP TEST KIT VTK2 20 CARTES A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more comp

DateMar 23, 2018
StatusTerminated
ReasonA shipment to Guadeloupe suffered a temperature excursions up to 47.8C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioMrieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
Root causeEnvironmental control
ActionThe firm, Biomerieux, sent an "Urgent product removal notice" letter on April 23 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Immediately acknowledge receipt (AR) of this FSCA. 2. Identify all countries and customers for which you are responsible that are impacted by this FSCA. 3. Define additional actions required (e.g. Translate customer letter, send to impacted countries/customers, etc.) 4. Distribute the customer letter to all customers that have received the products 5. Discard any impacte
Quantity15 boxes with GP TEST KIT VTK2 20 CARDS
DistributionInternational distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.