Recall Z-2140-2018 ยท DJR Enzyme Immunoassay, Methadone

Alere San Diego Inc. DBA Immunalysis Corporation

lmmunalysis EDDP Urine Control Set 1 The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tand

DateNov 28, 2017
StatusTerminated
ReasonThe High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.
Root causeUnder Investigation by firm
ActionOn November 28, 2017, the firm sent their customer notifications of the recall stating the following: CUSTOMER REQUIRED ACTION 1. If you have product in your possession: " Immediately discontinue use and/or sale of affected product lot and place in quarantine. " Complete the enclose Customer Verification Form, within 10 days, indicating the quantity of product you have discarded or will return to lmmunalysis. " FAX completed Customer Verification Form to 909-482-0850 or scan and email to rtimbang@immunalysis.com " Upon receipt of your Customer Verification Form, we will initiate shipment of
Quantity37 kits
DistributionUS Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151395Immunalysis EDDP Specific Urine Enzyme Immunoassay, Immunalysis EDDP Urine Calibrators, Immunalysis EDDP Urine Control S