ClearanceBrief
Recall Z-2175-2018 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Brilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray system

DateMar 30, 2018
StatusTerminated
ReasonThe patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.
Root causeDevice Design
ActionUrgent Field Safety Notices dated 3/28/18 were distributed to customers. The letters instructed customers to perform the following actions: Please review the following information with all members of your staff who need to be aware of the contents of this communication. It is important to understand the implications of this communication. Please retain a copy with the equipment Instruction for Use. As outlined in the Instructions for Use, always monitor the patient during all movements (manual or motorized) of the patient support. If you recognize an impending collision with the back shelt
Quantity2
DistributionWorldwide distribution including US states of CA, MD, NV, OH, TX, and WI, Afghanistan, Canada, Germany, Iraq, Italy, Korea, Kuwait, and Lebanon.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012009MX8000 IDT CT SYSTEM
Marconi Medical Systems, Inc., Sep 13, 2001