Recall Z-2178-2018 ยท DWE Tubing, Pump, Cardiopulmonary Bypass

Datascope Corporation

Maquet Cardiopulmonary Bypass Custom Tubing kit BO-TOP 38600 ADULT CPB, Part Number 701056035 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use only and is not to be re-sterilized by the user. The product consists of tubing and connectors with connected oxygenators, reservoirs, filters and other cardiopulmonary bypass components a

DateJul 27, 2017
StatusTerminated
ReasonThere is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.
Root causeNonconforming Material/Component
ActionOn 7/27/17, an Urgent Safety Alert was issued to customers of the Cardiovascular Procedure Kit with LN130B Over-pressure Safety Valves. The safety alert instructed customers to perform the following: Please examine your inventory immediately to determine if you have unused kits of the batch numbers listed. If so, please follow the recommended actions outlined in the attached Terumo Cardiovascular Systems Urgent Safety Alert. It is possible that custom tubing kits with the affected valves may have already been used without issue. Customers were instructed to test the valve prior to use. T
Quantity10
DistributionThe products were distributed to the following US states: IA and FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080592HLM TUBING SET WITH BIOLINE COATING
Maquet Cardiopulmonary, AG, Sep 11, 2008