Recall Z-1944-2018 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories
Olsen Medical
Symmetry Olsen bipolar forceps: (a1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 0.5mm Tips, REF 20-0311, QTY 5 (b1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 0.5mm Tips, REF 20-0311R, QTY 5 (c1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 1.0mm Tips, REF 20-0321, QTY 5 (d1) Insulated Irrigating Bayonet Forceps 6.25" (15.9cm) Smooth 1.5mm Tips, REF
| Date | May 11, 2018 |
| Status | Terminated |
| Reason | Products may encroach into the sterile pouch seal thereby increasing the risk of the seal being compromised. |
| Root cause | Package design/selection |
| Action | Symmetry initiated their recall via Urgent Medical Device Recall notices on 05/11/2018 sending faxes to accounts, as applicable; and, will also send the Notice via US certified mail, with return receipt requested, and will also send the Notice via e-mail to those customers for which it has a valid e-mail address (as applicable). All customers were asked to quarantine any products that reflect a breach of the seal area or any products that they do not wish to inspect immediately and promptly call (800) 251-3000 for a return material authorization (RMA) number to support the return of the impa |
| Quantity | 611,817 devices |
| Distribution | Worldwide Distribution |
| FDA record | Open on fda.gov |