Recall Z-1921-2018 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Smith & Nephew, Inc.

smith&nephew ANTHOLOGY POROUS HIGH OFFSET STEM, 12/14 TAPER, 7H SIZE, 32 MM x 112 MM, REF 71356107 Hip prosthesis.

DateMay 9, 2018 (initiated May 2, 2018)
StatusTerminated
ReasonThree lots of Anthology HO Porous size 7 stems were manufactured with an undersized taper length. The device taper length was incorrectly manufactured as 0.393mm opposed to 0.442 mm.
Root causeProcess control
ActionThe firm initiated the recall by letter on 05/02/2018. Consignees are asked to inspect inventory, locate and quarantine affected, unused devices. Affected product will be returned to Smith & Nephew. All consignees are asked to complete the Response Form, even if they do not have product to return. Distributors were directed to notify their customers of the field action and ensure that these actions are carried out. For further questions, please call (978) 749-1440
Quantity13 units
DistributionUS nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052792ANTHOLOGY HIP STEM
Smith & Nephew, Inc., Oct 7, 2005