ClearanceBrief
Recall Z-1939-2018 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

SOMATOM go.Now, Material Number 11061618 There is a potential for a software issue that may cause the need for necessary patient rescans.

DateApr 20, 2018
StatusTerminated
ReasonThere is a potential for a software issue that may cause the need for necessary patient rescans.
Root causeSoftware design
ActionThe firm, Siemens Healthineers, sent an "Customer Safety Advisory Notices" letter dated 4/20/2018 beginning on 4/20/18 to its customers. The letter described the product, problem and actions to be taken. The letters provided the following information to customers: How can the operator help to avoid a potential risk of the system? To avoid an unnecessary additional X-ray dose to patients please observe the following instructions: Topic 1: If a reconstruction cannot be started after an examination, please reboot the system and repeat the aborted reconstruction. If the error persists, please c
Quantity4 units in US
DistributionWorldwide Distribution: US (nationwide) including states of: AR, CA, CT, FL, GA, IL, IN, KS, KY, MO, ND, NY, OH, OK, SD, TN, TX, and VA; and countries of: Algeria, Australia, Austria, Bangladesh, Bolivia, Bosnia Herzog., Brazil, Cambodia, Chile, China, Columbia, Costa Rica, Germany, Greece, India, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K163296SOMATOM go.Up, SOMATOM go.Now
Siemens Medical Solutions USA, Inc., Mar 21, 2017