Recall Z-1954-2018 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical, Inc.

Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel.

DateApr 26, 2018
StatusTerminated
ReasonThe valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.
Root causeEmployee error
ActionOn April 27, 2018, the recalling firm contacted its consignees by telephone prior to sending out Urgent Medical Device Recall notifications via e-mail. Customers were instructed to remove affected devices from operation immediately. The firm provided replacement devices so customers could send their affected ventilator to the firm's repair center for update once the required parts were available. Customers were also asked to inform all affected users at each site. If you have any questions regarding this letter please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 1-80
Quantity8 units
DistributionDistributed to the states of CT, IL, MN, NV, SD, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072885CARINA SYSTEM, MODEL 8419350; CARINA BASIC DEVICE, MODEL 5704110
Draeger Medical AG & Co. KG, Apr 7, 2008