Recall Z-1953-2018 ยท DWS Instruments, Surgical, Cardiovascular

Genesee BioMedical, Inc.

Sternal Valve Retractors With Atrial and Aortic Rakes; Models SRCL-1, SRCL-AR, SRLP-1, SRLP-AR, ATR-1, ATR-AR, CSR-1 and CSR-AR. For use in cardiac surgery.

DateFeb 26, 2018
StatusOpen, Classified
ReasonFailure in a retractor blade due to corrosion. In the area of the blade where the bearing/button meets the blade body, corrosion can form, weakening the welded joint and may cause the bearing/button to break.
Root causeComponent design/selection
ActionThe customer communication letter, dated February 26, 2018, was mailed or emailed. The letter describes the concern and the user action. A copy of the revised Instructions for Use will be provided with the letter. All consignees in the US and outside of the US who have not been previously contacted will receive notification.
Quantity444 units
DistributionWorldwide distribution. US nationwide, South Korea, Australia, COSTA RICA, Taiwan, New Zealand, Italy, Canada, Netherlands, Norway, Sweden, Switzerland, Israel, Japan, Egypt, South Africa, United Kingdom
FDA recordOpen on fda.gov