Recall Z-2113-2018 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), infusion pump, 35700BAX2 Product Usage: The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixture of required patient therapy. The intended routes of administration cons

DateMay 8, 2018
StatusTerminated
ReasonThe firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets.
Root causeOther
ActionOn May 8, 2018, Baxter issued a Medical Device Correction letter to its customers. Customers were advised of the issue and instructed to do the following: 1. The SIGMA SPECTRUM Infusion Pumps can continue to be safely used with the Baxter-provided 12.5-inch hanger. 2. The updated Instructions for Use will be made available by May 31, 2018. At that time, you may access the updated IFU online at the Baxter Global Technical E-Service Center at URL: https://service.baxter.com/tsportal/#/login. For additional questions and concerns about this issue please contact Baxter Healthcare Medina at 800-35
Quantity122889
DistributionWorldwide Distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133801BAXTER SIGMA SPECTRUM INFUSION PUMP WITH MASTER DRUG LIBRARY
Baxter Healthcare Corporation, May 7, 2014