Recall Z-2100-2018 ยท OVD Intervertebral Fusion Device With Integrated Fixation, Lumbar

VGI Medical, LLC

VGI MEDICAL VerteLP LATERAL INTERBODY FUSION instrument kit

DateApr 9, 2018
StatusTerminated
ReasonIFU contained an automated cleaning process for the VerteLP Instrument Tray that was not validated
Root causeProcess design
ActionThe firm notified their direct consignees by letter on 04/09/2018 and followed with an e-mail on 04/11/2018. The letter directed the direct consignees (distributors) to instruct their accounts (user) to discontinue the automated cleaning process and to provide users with the new IFUs.
Quantity6 units
DistributionDistributed to MI, FL, and NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151312VerteLP Interbody Fusion Device
Vgi, LLC, Jan 13, 2016