ClearanceBrief
Recall Z-2269-2018 ยท FET Endoscopic Video Imaging System/Component, Gastroenterology-Urology

Olympus Corporation of the Americas

Olympus JF-140F duodenoscopes (Model NumberJF-140F)

DateMay 11, 2018
StatusTerminated
ReasonIssuance of validated, new reprocessing procedures. This corrective action is being undertaken as a result of ongoing manufacturer and FDA activities relating to reported patient infections associated with duodenoscopes. The new cleaning procedure requires additional recess flushing and forceps elevator raising/lowering steps during precleaning, manual cleaning and manual disinfection.
Root causeUnder Investigation by firm
ActionFirm is extending recall to the user level Letters will be sent on May, 11, 2018 via overnight mail to customers Follow-up phone calls to consignees to offer training on new reprocessing instructions No action can be taken with hospitals that are no longer in business
Quantity994
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K853585OLYMPUS EVS-ENDOSCOPIC VIDEO IMAGE & DATA SYS
Olympus Corp., Jan 21, 1986