ClearanceBrief
Recall Z-1806-2018 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer Biomet, Inc.

Persona Stemmed Tibial Provisional Size H Right, Item Number 42532108302 For use during orthopaedic surgery for implantation of a prosthesis.

DateAug 7, 2017
StatusTerminated
ReasonThere is a potential for intermittent cracks in the raw material batch used to produce the affected products. The cleanliness of the affected products could be compromised if cracks penetrate the surface of the instrument as a result of the raw material issue.
Root causeUnder Investigation by firm
ActionThe firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL (REMOVAL) LOT SPECIFIC" letter dated 8/7/17 to its Customers. The letter described the product, problem and actions to be taken. The Distributors were instructed to: 1. Review this notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form a. Return a digital copy to corporatequality.postmarket@zimmerbiomet.com within three (3) days. 4.
Quantity842 products in total
DistributionWorldwide Distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MT, NC, NJ, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI, and Puerto Rico; and to countries of: Australia, Canada, Hong Kong, India, Japan, Korea, Malaysia, Netherlands, and Singap
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113369ZIMMER PERSONA KNEE SYSTEM
Zimmer, Inc., Mar 27, 2012