Recall Z-1705-2018 ยท JEY Plate, Bone

Zimmer Biomet, Inc.

Lactosorb RapidFlap, bone plate, Model No. 915-0020 The RapidFlap LS Cranial Fixation System is indicated for use in pediatric craniotomy flap fixation.

DateMay 7, 2018 (initiated Apr 5, 2018)
StatusTerminated
ReasonThe recalling firm has confirmed that the Outer Plate component exhibits an excessive chamfer on the threading after deburring operations. This excessive chamfer results in non-conforming product where the threads of the outer plate component have limited to no engagement with the post component.
Root causeProcess control
ActionZimmer Biomet sent an Urgent Medical Device Recall letter dated April 5, 2018, to Distributors, Sales Representatives, and Distribution Managers. A separate recall letter was sent to Risk Managers on the same date. (Updated 3/20/19) On February 12, 2019, the firm issued notification of recall expansion. Further affected lots had been identified as part of the recall, and affected customers were notified. (/Update) Consignees were informed of the product issue and risk of non-functioning device and delays in surgery. All parties were instructed to review the notice, locate and quarantine aff
Quantity13175
Distributionworldwide Distribution - US Distribution to the states of : CA, FL, MO, NC, SC, TX, and WI., and to the countries of : Argentina, China, Colombia, Germany, Japan, Netherlands, South Africa, Thailand, and Vietnam.