ClearanceBrief
Recall Z-1908-2018 ยท LHI Set, I.V. Fluid Transfer

Baxter Healthcare Corporation

Baxter EXACTAMIX Inlet, Non-Vented, High-Volume Inlet, REF H938173 Product Usage: Exacta-Mix 2400 Compounding System Administration set is a disposable component of a compounding device used in the pharmacy to compound multiple source ingredients into one final solution.

DateApr 25, 2018
StatusTerminated
ReasonSpecific lots of EXACTAMIX Inlets listed below due to incorrectly performed over-labeling. The inlets were over-labeled with a blank label, or with a label containing an extended expiration date. If the product is used beyond the expiration date,
Root causeLabeling Change Control
ActionThe firm initiated the recall by letter on 04/25/2018. The letter identified the affected product, problem and actions to be taken. the firm requested the return of the product. Distributors were directed to conduct a sub-recall, requesting their customers return the product to the entity from whom they received the product.
Quantity98700 units
DistributionWorldwide Distribution - US Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K002705EXACTA-MIX 2400 COMPOUNDING SYSTEM, MODEL EXACTA-MIX 2400
Baxa Corp., Mar 28, 2001