Recall Z-1937-2018 ยท FTD Lamp, Surgical

Maquet SAS

VOLISTA StandOP (model 600) surgical lights shipped from October 2012 until June 2017 equipped with keypads with Ref. ARD568801560; Part numbers: 568811901, VLT600SF AIM STP; 568811911, VLT600SF AIM STP; 568811961, VLT600DF AIM STP; 568821910, VLT600SF STP QL; 568821911, VLT600SF AIM STP QL; 568821960, VLT600DF STP QL; 568821961, VLT600DF AIM STP QL VOLISTA surgical lights are intended to be use

DateApr 19, 2018
StatusTerminated
ReasonPotential keypad failure on VOLISTA StandOP Surgical Lights.
Root causeDevice Design
ActionThe firm, Getting, sent "Urgent Medical Device Recall Notice" letters was issued April 19, 2018 and April 24, 2018 via 2 Day Fedex with customer signature required to allow for tracking of delivery. The letter includes a customer response form requesting feedback including if the device is still in use at the facility via FAX BACK TO: 1-585-486-5605 or EMAIL TO: VOLISTA_StandOP_Keypad_FA.SW-US@getinge.com. The letter instructs the customer to contact Getinge Service to schedule a visit and replacement by a trained Getinge technician at no charge to the customer. In addition Getinge Service wi
Quantity187
DistributionWorldwide Distribution: US (nationwide) and countries of: (AE, AR, AT, AU, BD, BE, BO, BR, CA, CH, CL, CM, CN, CO, CZ, DE, DK, EC, EE, EG, ES, Fl, FR, GB, GH, HK, HR, HU, IE, IN, IQ, IR, IT, JO, JP, KE, KR, LB, LU, LV, MQ, MX, MY, NA, NG, NL, OM, PA, PK, PL, PT, PY, RO, RS, SA, SE, SG, SI, SK, TH, T
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K130513VOLISTA 400; VOLISTA 600
Maquet S.A.S., May 8, 2013