Recall Z-1802-2018 ยท DQY Catheter, Percutaneous
Bard Peripheral Vascular Inc
BARD PERIPHERAL VASCULAR SEEKER(R) Crossing Support Catheter, 0.035", 150 cm, REF SK15035M. Intended to support a guidewire during access into the vasculature.
| Date | May 11, 2018 (initiated Apr 16, 2018) |
| Status | Terminated |
| Reason | Device is at risk of having detectable levels of bacterial endotoxin present on the packaging hoop that the catheter is packaged in. The bacterial endotoxin levels for the catheters are within acceptable levels; however, if bacterial endotoxins are present on the packaging hoops it may have the potential for cross-contamination in a clinical setting. |
| Root cause | Process control |
| Action | The firm initiated the recall via an Urgent Medical Device Recall Notification letter on 04/16/2018. The letter identified the affected device, reason for the recall, and required actions. Customers are asked to check their inventory for and remove the affected product code and lot number combinations. The Recall and Effectiveness Check Form should be completed and returned even if affected product has been used. The firm requested the return of the recalled product; and will provide a replacement after all information has been verified and product returned. The completed form can be emai |
| Quantity | 140 units |
| Distribution | Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K103367 | SEEKER CROSSING SUPPORT CATHETER Bard Peripheral Vascular, Inc., Dec 8, 2010 |