Recall Z-1894-2018 ยท KQO Automated Urinalysis System

Beckman Coulter Inc.

iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry System Computerless, Catalog #700-7177-001. The iChemVELOCITY automated urine chemistry system is an in-vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, a

DateApr 27, 2018 (initiated Mar 29, 2018)
StatusTerminated
ReasonThe Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.
Root causeProcess control
ActionThe recalling firm, Beckman Coulter/IRIS, sent an "URGENT MEDICAL DEVICE RECALL" letter dated dated 3/26/2018 via email on 3/29/2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: -Follow the on-board strip stability claims per Table 1 to determine the frequency at which the strips need to be changed. -Whenever the Strip Provider Module is accessed (i.e., opened or when loading strips), remove and discard the remaining on-board strips and desiccant before replenishing. -It is suggested that for the tests p
Quantity6 devices
DistributionUS Distribution to states of: AL, IA, GA, and FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101852CHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM, STRIPS AND CALCHEK KIT
Iris International Inc., A Division of Iris Iris I, Mar 23, 2011