Recall Z-1920-2018 ยท DZE Implant, Endosseous, Root-Form

Implant Direct Sybron Manufacturing, LLC

ImplantDirect Legacy 4 Implant, 5.7mmD x 5.7mmD Platform, Part 885708, Rx, Sterile. The incorrect cap label shows part 855708, 5.7mmD Platform, 8mmL.

DateNov 15, 2016
StatusTerminated
ReasonThe vial cap was mislabeled with the incorrect part number of 855708 but the product description on the cap was correct. The main vial label was correctly labeled.
Root causeLabeling mix-ups
ActionLetters dated 11/14/2016 were issued via FedEx overnight on 11/15/2016.
Quantity18 units
DistributionDistribution was made to FL, GA, MD, MI, NC, NY, OR, TX, UT and VA. There was no foreign/military/government distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090234SPECTRA SYSTEM DENTAL IMPLANTS 2008
Implant Direct, LLC, May 15, 2009