Recall Z-1813-2018 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Draegar Medical Systems, Inc.

Infinity Acute Care System (lACS) Monitoring Solution; Catalog Numbers:MS20401, MS20724, MS22956, MS25510, MS25520, MS25643, MS26196, MS26372, MS31818; UDI Information: 4049098054454, 4049098054447, 4049098054409, 4049098009799, 4049098009751. Multi-parameter, physiologic patient monitoring of adult, pediatric and neonatal patients in environments where patient care is provided by trained healt

DateMar 28, 2018
StatusTerminated
ReasonSoftware anomaly resulting in the loss of patient settings and stored patient data.
Root causeUnder Investigation by firm
ActionOn March 28, 2018, Drager Medical Systems Inc. mailed Urgent Medical Device Recall Notifications to affected customers. The following information was included in the notification: 1) Do not use Infinity MCable Mainstream CO2 sensors revision 16 with M540 Patient Monitor, software versions VG4.1 or lower. Previous revisions of the CO2 sensor may be used. 2) We are in process of resolving this product issue. If you choose to upgrade your system(s), your Drger service representative will schedule a service visit and upgrade your M540 patient monitor(s) once the updated software is availab
Quantity25, 629
DistributionWorldwide Distribution - USA (nationwide) and to the countries of : Angola, Argentina, Australia Austria Azerbaijan Bahrain Belgium Bolivia Brazil Brunei Dar-es-S Canada Chile China Colombia Costa Rica Czech Republic Denmark Egypt Finland France Germany Ghana Greece Hong Kong Hungary India Indonesia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113798INFINITY ACUTE CARE SYSTEM, INFINITY MCABLE- MASIMO RAINBOW SET
Draeger Medical Systems, Inc., Mar 22, 2012