Recall Z-1893-2018 ยท LJX Test, Urine Leukocyte

Beckman Coulter Inc.

iChem VELOCITY Urine Chemistry Strips, REF 800-7212 Product Usage: The iChem VELOCITY automated urine chemistry system is an in vitro diagnostic deice used to automate the urine chemistry analysis profile using iChem VELOCITY Urine Chemistry Strips. The iChem VELOCITY can be used as a stand alone-system, as well as in an iQ200 Series system, a configuration given the proprietary name iRICELL as i

DateDec 4, 2017
StatusTerminated
ReasonA subset of Lot 7212154 A of the iChem VELOCITY Urine Chemistry strips has an incorrect pad placed in the location of the leukocytes pad. This will result in control CB failure for leukocytes and may generate false negative erroneous results for leukocytes.
Root causeMixed-up of materials/components
ActionThe firm initiated the recall by letter on 12/04/2017. The letter identified the affected product, problem and the actions to be taken. The firm requested destruction of the recalled product.
Quantity14340 units
DistributionWorldwide Distribution and US Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101852CHEM VELOCITY AUTOMATED URINE CHEMISTRY SYSTEM, STRIPS AND CALCHEK KIT