ClearanceBrief
Recall Z-1759-2018 ยท GAD Retractor

Synthes (USA) Products LLC

Retractor f/Sciatic Nerve Long; Part number: 03.100.014; UDI: 10886982070630

DateApr 2, 2018
StatusTerminated
ReasonThere is a potential for microspores to form on the hollow handle of the Sciatic Nerve Retractor. The pores may increase in size, allowing fluid to enter the hollow handle.
Root causeUnder Investigation by firm
ActionOn April 2, 2018 Synthes (USA) LLC mailed an Urgent Product Recall notification letter to affected customers. Customers were instructed to: Review current inventory to identify and quarantine all affected products; Complete the Verification section of the Return Receipt form; Return affected product; Forward the notification to anyone in your facility that needs to be informed; If affected products have been forwarded to another facility, contact that facility to arrange return; Keep a copy of the notification. Customers with questions may call (610)719-5450 or contact your Depuy Synthese Sa
Quantity196
DistributionUS Nationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.