Recall Z-2123-2018 ยท HRS Plate, Fixation, Bone

Orthofix, Inc

Non-Locking Screw, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Number): a)D 3.2MM X 10MM (002-D-32010) b) D 3.2MM X 12MM (002-D-32012) c) D 3.2MM X 14MM (002-D-32014) d) D 3.2MM X 16MM (002-D-32016) e) D 3.2MM X 18MM (002-D-32018) f) D 3.2MM X 20MM (002-D-32020) g) D 3.2MM X 22MM (002-D-32022) h) D 3.2MM X 24MM (002-D-32024) i) D 3.2MM X 26MM (002-D-32026) j) D 3.2MM X 28MM (002-D

DateJun 7, 2018 (initiated Apr 6, 2018)
StatusTerminated
ReasonFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Root causeDevice Design
ActionOn 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
Quantity1206
DistributionU.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140876VIEW PLATING SYSTEM,REDUCE FRACTURE PLATING SYSTEM,BONE SCREW
Icon Orthopedic Concepts, LLC Dba Edge Orthopaedic, Jul 22, 2014