Recall Z-1896-2018 ยท MNL Accessories, Cleaning Brushes, For Endoscope
Madison Polymeric Engineering
First Step Endoscopic Cleaning Pad, 200ml, Flexible Endoscope Bedside Pre-Clean Kit, Catalogue Number EP-4
| Date | Mar 21, 2018 |
| Status | Terminated |
| Reason | The products exhibit an unpleasant odor due to microbial contamination with Yarrowia lipolytica. |
| Root cause | Under Investigation by firm |
| Action | Notification letters were distributed on 3/21/18. The letters instructed customers to perform the following: 1. Immediately quarantine, and discontinue the use or distribution of First Step EP-3, EP-4, EP-6, 4" Button Brush EP-4B; Travel Kit TK200(W); and Draco Enzymatic Deep-Cleaning Pad HY0305Z with the referenced lot numbers specified in Table1. 2. Determine if any stock of the with the reference lot numbers specified in Table 1 remains in your inventory. We ask that you discard the product at your facility and document the quantity discarded on the form. Otherwise, we can issue you a R |
| Quantity | 1422 cases/99540 kits |
| Distribution | The products were distributed to the following US states: AK, AL, AR, CA, CT, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, and WI. The products were also distributed in Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.