ClearanceBrief
Recall Z-2118-2018 ยท HRS Plate, Fixation, Bone

Orthofix, Inc

Rival Reduce Kit Plates, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Part Number: 000-K-012C

DateJun 7, 2018 (initiated Apr 6, 2018)
StatusTerminated
ReasonFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
Root causeDevice Design
ActionOn 04/06/2018, Orthofix Inc. (U.S. Agent), sent written notification of the recall to distributors and hospitals via e-mail and/or FedEx overnight courier.
Quantity50
DistributionU.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142135REDUCE FRACTURE PLATING SYSTEM LINE EXTENSION
Icon Orthopedic Concepts, LLC Dba Edge Orthopaedic, Aug 28, 2014