ClearanceBrief
Recall Z-1888-2018 ยท GCJ Laparoscope, General & Plastic Surgery

Ethicon Endo-Surgery Inc

5mm Endopath Xcel with Optiview Technology, Universal Trocar with Stability Sleeve, 5mm, 100mm shaft length, 2CB5LT

DateMar 26, 2018
StatusTerminated
ReasonUse of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal.
Root causeProcess control
ActionThe firm notified customers and distributors by letter on March 26, 2018, and March 27, 2018, respectively. Customers were informed of the issue and potential risk of prolonged surgery or introduction of minute particles. Customers were instructed to: 1. Examine your inventory immediately to determine if you have any products subject to this recall on hand and quarantine such product(s). 2. Remove the products subject to this recall from your inventory and communicate the issue to all relevant operating room or materials management personnel, or anyone else in your facility who needs to be i
Quantity95233
DistributionNationwide domestic distribution, including PR. International distribution to Saudi Arabia, United Arab Emirates, Belgium, Singapore, Taiwan, South Korea, Canada, Chile, Mexico, Japan, and Argentina.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032676ENDOPATH BLADELESS TROCAR, ENDOPATH BLUNT TIP TROCAR, ENDOPATH DILATING TIP TROCAR
Ethicon Endo-Surgery, Inc., Oct 30, 2003