Recall Z-1888-2018 ยท GCJ Laparoscope, General & Plastic Surgery
Ethicon Endo-Surgery Inc
5mm Endopath Xcel with Optiview Technology, Universal Trocar with Stability Sleeve, 5mm, 100mm shaft length, 2CB5LT
| Date | Mar 26, 2018 |
| Status | Terminated |
| Reason | Use of the product with insufficient lubrication may prolong surgery due to trocar replacement. In addition, insufficient lubrication could result in a remote chance of introducing a minute particle if repeated insertions of the obturator or laparoscopic device damage the seal. |
| Root cause | Process control |
| Action | The firm notified customers and distributors by letter on March 26, 2018, and March 27, 2018, respectively. Customers were informed of the issue and potential risk of prolonged surgery or introduction of minute particles. Customers were instructed to: 1. Examine your inventory immediately to determine if you have any products subject to this recall on hand and quarantine such product(s). 2. Remove the products subject to this recall from your inventory and communicate the issue to all relevant operating room or materials management personnel, or anyone else in your facility who needs to be i |
| Quantity | 95233 |
| Distribution | Nationwide domestic distribution, including PR. International distribution to Saudi Arabia, United Arab Emirates, Belgium, Singapore, Taiwan, South Korea, Canada, Chile, Mexico, Japan, and Argentina. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K032676 | ENDOPATH BLADELESS TROCAR, ENDOPATH BLUNT TIP TROCAR, ENDOPATH DILATING TIP TROCAR Ethicon Endo-Surgery, Inc., Oct 30, 2003 |