Recall Z-1744-2018 ยท LCO Platelet Factor 4 Radioimmunoassay

Hyphen BioMed

ZYMUTEST HIA IgGAM; REF Number: RK040D; UDI: 03663537018770 a qualitative screening assay intended for the global detection of heparin- dependent antibodies, whether the isotype is: IgG, IgM and IgA, in human plasma, by clinical laboratories.

DateFeb 5, 2018
StatusTerminated
ReasonIncorrect package insert indicating use of the kits with plasma, serum, and other biological fluids. The kits should only be used with plasma.
Root causePackaging process control
ActionOn March 6, 3018 Aniara Diagnostica began emailing Urgent Medical Device Recall notifications to affected customers. Customers are instructed to: Use the US Instructions for Use and Return the completed Acknowledgement Awareness Form. Customers with questions may contact Aniara Diagnostica via e-mail at Sandra@aniara.com or by phone at (866)783-3797 ext. 110.
Quantity281
DistributionOH, MD, NC, NY, NH, Canada, Sweden
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071255ZYMUTEST HIA IGGAM, ZYMUTEST IGG, MODEL RK040D, RK040A
Hyphen Biomed, Sep 7, 2007