ClearanceBrief
Recall Z-2438-2018 ยท CDZ Radioimmunoassay, Testosterones And Dihydrotestosterone

Siemens Healthcare Diagnostics, Inc.

Dimension Vista MULTI 2 SDIL, KD694, SMN# 10483586

DateMar 21, 2018
StatusTerminated
ReasonThe diluent may have an incomplete slit on the septum in the cap of the vial. A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.
Root causeUnder Investigation by firm
ActionNotification letters were distributed 3/21/18. The letters instructed customers to: Inspect stock, discontinue use of and discard the Dimension Vista Diluent lots listed in Table 1. Please review this letter with your Medical Director. Review your inventory of these products to determine your laboratorys replacement needs and to provide information to Siemens for reporting to the authorities. Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. If you have received any complaints of illness or adverse events associated with the produc
Quantity1195
DistributionThe products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The products were distributed to
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151529Dimension Vista LOCI Total Testosterone Flex reagent cartridge, Dimension Vista Testosterone Calibrator
Siemens Healthcare Diagnostics, Inc., Feb 11, 2016