ClearanceBrief
Recall Z-1604-2018 ยท IXR Table, Radiographic, Tilting

Philips Medical Systems Nederlands

708033 URODiagnost As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

DateMay 3, 2018 (initiated Mar 26, 2018)
StatusTerminated
ReasonThe fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
Root causeNonconforming Material/Component
ActionThe firm, BU IGT Systems, sent an "URGENT- Field Safety Notice" letter dated 3/23/2018 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to stop using the system and call your Philips representative if the following situations occurs: -the table height or tilt movement is not working. - you notice a cracking or snapping sound during system operation. - you experience a blockade in the table height or tilt movement during system movements. -the system is having an unexpected collision or the system had an unexpected collision in
Quantity20
DistributionWorldwide Distribution: US (Nationwide) and countries of: Algeria, Argentina, Australia, Austria, Belgium, Canada, China, Colombia, Czech Republic, Denmark, Egypt, Estonia, Farce Islands, Finland, France, Germany, Hong Kong, Indonesia, Ireland, Israel, Italy, Jersey, Jordan, Korea, Republic of, Latv
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911098PHILIPS URO DIAGNOST MRF
Philips Medical Systems, Inc., Jul 3, 1991