Recall Z-1589-2018 ยท OWB Interventional Fluoroscopic X-Ray System

Philips Electronics North America Corporation

Allura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, eg peripheral, cerebral, thor

DateMar 14, 2018
StatusTerminated
ReasonAfter continuous operation for more than one and a half days, the image on the large screen monitor may freeze for approximately 15 seconds after which the system will restore itself.
Root causeSoftware design
ActionField Safety Notices were distributed on 3/14/18. The notices instructed customers: To prevent occurrence of the issue customers shall restart the system at least once a day. Customer shall ensure that all staff with access to the affected systems are informed of the contents of this Field Safety Notice. A copy of this Field Safety Notice shall be placed together with the documentation of the system until the system has been corrected by Philips. The problems will be resolved by a software update, which is expected to be available by the second half of 2018. You will be notified by you
Quantity682
DistributionWorldwide Distribution in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The product was distributed to the following countries: Argentina, Australia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133292ALLURA XPER FD; OR TABLE
Philips Medical Systems Nederland B.V., Mar 5, 2014